Seta – Pensky-Martens Flash Point Tester – PM-93
The Seta PM-93 is a fully automated Pensky-Martens Closed Cup Flash Point Tester which combines strict method conformance with the latest control technology and safety systems to provide the next generation of flash point instruments.
An intuitive graphic user interface and 8.4” SVGA colour touch screen operation provide unparalleled ease of use with real time display of test progress. Test profiles, operator names, test methods and storage of unlimited results are provided through the embedded operating system. Test results can be emailed, saved to a memory stick, output to LIMS or produced in hard copy with the printer option.
Technical Specifications
Ignition system | Electric or gas |
Flash detection | Thermal |
---|---|
Cooling | Forced air (integral fan) |
Heating Rate | Fast heating rate (>10°C/min) and standard; 5.5°C/min; 3°C/min; 1.3°C/min; 1°C/min |
Temperature range | Ambient +5°C to 400°C |
Barometric pressure correction | Automatic correction with built-in pressure sensor |
Internal Memory | 30 programmable test profiles, operator names, test methods and unlimited tests results |
Fire Detection | Thermal with optional integral Fire Extinguisher |
Dimensions (H xW x D) / weight | 38.5 x 24 x 50 cm / 25 kg |
Voltage | 230 Vac +/-15% or 110 Vac +/-15%. Frequency 50/60 Hz. Auto-sensing – no selector switch required. |
Key Features
PM-93 – Pensky-Martens Flash Point Tester
- New design, latest enhanced operation
- Fully conforms to test methods with 4 heating rates available
- Simple ‘3-step to test’ operation
- Unique patented ‘SafeFlash’ flame extinguisher system
- Large 8.4” Resistive touch screen menu
- Single action lifting arm operation
- Pre-dip safety mode
- Programmable with 30 test profiles, test methods & sample identities
- Test memory = 2000 results
- Gas or electronic ignition using proven and long lasting Seta ‘Ignite’ technology
- Compact footprint
Testing Methods And Details
BS EN ISO 2719, BS 2000-34 | Methods of test for petroleum and its products. Flash point by Pensky-Martens closed tester |
ASTM D93 | Standard test methods for flash-point by Pensky-Martens closed cup tester |
---|---|
IP 34 | Determination of flash point – Pensky – Martens closed cup method (ISO 2719:2002) |
Pensky-Martens Flash Point Testing by PM-93
Pensky-Martens Flash Point Tester for ASTM D93 | IP 34 | ISO 2719 by Seta PM-93
Key Industries
- Environmental
- Oil & Gas
- Pharmaceutical
Brand
Stanhope-Seta design and manufacture quality control instruments used to measure the physical characteristics that determine product quality and consistency. The ‘Seta’ range includes test equipment for the hydrocarbons industry as well as many other laboratory test applications for quality control, calibration and analysis.